SYSTEM SETUP, Flag System
11-30 Sysmex SF-3000 Operator's Manual -- Revised September 1995
NOTE: RBC IP messages are displayed in the order of their Item Nos.
Messages with smaller Item Nos. are displayed first, followed by
those with larger Item Nos.
(4) Press [QUIT] key to exit the program. The confirmation message will appear.
Press the appropriate key [Continue], [Set] or [Cancel].
Continue : To return to RBC setting screen so that setting can be continued.
Set : To save the new settings. Program ends.
Cancel : To cancel the new settings. Program ends.
4.3 PLT
Use this program to set IP message limits for PLT.
To set abnormal limits for PLT:
(1) Press [PLT] key in the Flag System submenu.
The PLT setting screen containing the current settings for the various IP messages
and numeric keys will appear.
(2) Use the [] and [] keys to select the desired abnormal message and setting item.
(3) Use the numeric keys to input the new limit value. Press [ENTER] to input the
values and move down to the next item.
WBC RBC PLT Return
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
ABNORMAL
COMMENT
PLT
Thrombocytopenia
3
ITEM
ENTER
QUIT
C
78 9
456
123
0
60
PLT
PLT [X10 /uL]60
PLT
600
3
[X10 /uL]
SET VALUE
Thrombocytosis
Figure 11-23: PLT Flag System Setting Window