SYSTEM SETUP
Sysmex SF-3000 Operator's Manual -- Revised September 1995 11-31
Table 11-4 : PLT Abnormal IP Messages (The Factory Settings)
Item No. Message Message occurs when
2 Thrombocytopenia PLT < 60 [x 10
3
/µL]
3 Thrombocytosis PLT > 600 [x 10
3
/µL]
NOTE: The factory settings are given in Table 11-4.
NOTE: PLT IP messages are displayed in the order of their Item Nos.
Messages with smaller Item Nos. are displayed first, followed by
those with larger Item Nos.
(4) Press [QUIT] key to exit the program. The confirmation message will appear.
Press the appropriate key [Continue], [Set] or [Cancel].
Continue : To return to PLT setting screen so that setting can be continued.
Set : To save the new settings. Program ends.
Cancel : To cancel the new settings. Program ends.