TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-35
7 . Manual CP Unit Errors
Set Piercer Lid
Phenomena Piercer lid is open when the Start Switch is pressed.
Probable Cause Sensor switch to detect closed lid is defective.
Corrective Action 1) Make sure to close the piercer lid.
2) If the error persists, contact your Sysmex service representative.
Set Manual Piercer Cover
Phenomena Manual CP cover is not installed when the Start Switch is pressed.
Probable Cause Unit cover detection sensor is defective.
Corrective Action 1) Make sure to install the Manual CP cover.
2) If the error persists, contact your Sysmex service representative.