MAINTENANCE & SUPPLIES
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-7
4. WEEKLY MAINTENANCE
4.1 Clean SRV Tray
Clean the tray under the Sample Rotor Valve approximately once a week, or every 600
samples, as described below.
WARNING! The SRV and Tray should be considered potentially biohazardous.
Do not handle the SRV or the Tray without proper protective
equipment. Wear rubber gloves when cleaning. Wash your hands
with an antiseptic solution after completing the procedure.
(1) Turn OFF the power to the Main Unit. Wait at least 30 seconds to release the
pressure and vacuum accumulated in the pneumatic lines.
(2) Open the front cover of the Main Unit and set the stop bar in position so that the
cover will not close accidentally.
(3) Remove the SRV Tray by pulling it to the left. See Figure 5-4-1.
CAUTION: When removing the SRV tray, do not loosen the thumbscrew fixing
the Manual Aspiration Pipette on the unit.
(4) Rinse the tray with water and wipe dry. See Figure 5-4-2.
SRV Tray
Thumbscrew
Tray
Figure 5-4-1: Remove SRV Tray Figure 5-4-2: Rinse SRV Tray
(View from the instrument
left side)