SYSTEM SETUP
Sysmex SF-3000 Operator's Manual -- Revised September 1995 11-1
1. INTRODUCTION
System setup of the SF-3000 will be performed by the Sysmex Service Representative
when the instrument is installed. However, the settings can be changed according to
each laboratory's needs by using the Setting Programs explained in this chapter. This
chapter consists of the following sections.
Auto Management
Explains setting conditions for the initiation of Auto Output (which will direct
automatic data transmission and/or printout), Auto Erase (which allows automatic
data erasing when the memory is full), and Auto Validation (which validates the
previously specified data allowing automatic output of results).
Data Criteria
Explains setting upper and lower limits for abnormal patient data (marking of
abnormal data with flags).
• Mark Limits
Explains setting upper and lower limit ranges for abnormal patient data.
• Critical Limits
Explains setting upper and lower limits for critical data.
• Sampler Stop Limits
Explains setting upper and lower limits for stopping the sampler's operation.
• Sampler Stop Conditions
Explains setting the various conditions for stopping the sampler's operation.
Flag System
Explains setting criteria for flagging of WBC, RBC and PLT abnormal IP
messages.
Output Condition
Explains setting of printer output conditions, print formats and interface conditions
for transmission of data to each type of peripheral device.