SYSTEM SETUP
11-48 Sysmex SF-3000 Operator's Manual -- Revised March 1996
6.4 Timer Mode
After sample analysis is complete and no key operation is performed within the set period
of time, the timer can automatically switch off the power to the Pneumatic Unit. Use this
program to set the time.
To set timer mode:
(1) Press [Timer Mode] key in the General Set Up submenu. The Input Password
window will appear.
(2) Enter the password and press [ENTER] key.
The Timer Mode setting screen containing the current setting and numeric keys will
appear.
(3) Input the period of time before the pneumatic unit is turned off. The available
setting time is "in minutes" ranged from 1 to 15.
(The factory setting is "10" minutes.)
10
Delay Time
Delay Time
Date/
Time
Units
Normal
Range
Date
Format
Timer
Mode
Password
More
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
TIMER MODE
10
[min]
QUIT
ENTER
C
3
9
6
0
1
7
4
2
8
5
Return
Figure 11-40: Timer Mode Setting Window
(4) Press [QUIT] key to exit the program. The confirmation message will appear.
Press the appropriate key [Continue], [Set] or [Cancel].
Continue : To return to Timer Mode setting screen so that setting can be
continued.
Set : To save the new settings. Program ends.
Cancel : To cancel the new settings. Program ends.