EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #177 background imageLoading...
Page #177 background image
MAINTENANCE & SUPPLIES, Daily
5-6 Sysmex SF-3000 Operator's Manual -- Revised October 1995
3.2 Remove Fluid from Pneumatic Unit Trap Chamber
At the end of each day's analyses, check the fluid level in the pneumatic unit trap
chamber. Discard any fluid that has collected.
WARNING! The contents in the Trap Chamber should be considered potentially
biohazardous. Do not handle the Trap Chamber without proper
protective equipment. Wear rubber gloves when discarding the
fluid. Wash your hands with an antiseptic solution after completing
the procedure.
(1) Turn OFF the power to the Main Unit. Wait at least 30 seconds to release the
pressure and vacuum accumulated in the pneumatic lines.
(2) Verify that the pneumatic gauges on the Pneumatic Unit front show the zero level.
(3) Turn the trap chamber counterclockwise and remove. See Figure below.
Trap Chamber
Loosen
Figure 5-3-4: Trap Chamber (Front of the Pneumatic Unit)
(4) Discard the fluid; then reattach the chamber.
CAUTION: If fluid collects every day, contact a Sysmex service representative.
(5) Keep your log on the maintenance check list.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals