MAINTENANCE & SUPPLIES, Daily, Shutdown
Sysmex SF-3000 Operator's Manual -- Revised October 1995 5-5
(3) Position the CELLCLEAN [or 5%
filtered sodium hypochlorite (bleach)]
into the manual aspiration pipette as
shown in Figure 5-3-2; then press the
Start Switch. Do not remove the
CELLCLEAN while the READY LED is
blinking as CELLCLEAN is being
aspirated.
(4) After the READY LED turns off and the
alarm beeps three times, remove the
CELLCLEAN.
(5) The shutdown sequence will start.
During shutdown sequence, "Shutdown
in Progress" is displayed on the LCD.
Start Switch
Ready LED
CELLCLEAN
Figure 5-3-2:
Introduce CELLCLEAN
(6) After the shutdown sequence is completed, a continuous beep will sound and the
Shutdown Complete screen will appear.
Restart
Turn OFF power to shut down.
Press Restart to continue analyses.
X
M DP:
Sysmex
Ready
Sampler
Shutdown Next No.
HC GP DP SU
Shutdown
Figure 5-3-3: Shutdown Complete Screen
(7) If you are going to shutdown the instrument, turn the Main Unit power OFF. Turn
the printers OFF, if attached.
NOTE: If you are going to continue to operate, perform a background check
by pressing the [Restart] key displayed on the LCD.