TECHNICAL INFORMATION, Serial Interface
C-4 Sysmex SF-3000 Operator's Manual -- Revised September 1995
1.7 Software
1. Code
ASCII codes are used in this interface.
2. General Function
FUNCTION DESCRIPTION
1. Analysis data output
Auto Out -- The SF-3000 automatically sends out analysis
data after each analysis has been completed.
Stored Data -- The SF-3000 sends out data from the stored
data in a batch.
2. Inquiry (output) and settings from
host computer (input)
When SF-3000 inquires order information by the Rack No.
and Tube Position No., the host computer gives order
information and the sample ID number for each sample in
the rack to SF-3000.
Table C-4: General Functions
3. Framing of Text
STX (02 in hexadecimal code) is sent prior to data and ETX (03 in hexadecimal
code) is sent following data. The text length is within 255 bytes.
S
T
X
E
T
X
Order of Transmission
Figure C-3: Order of Transmission