CALIBRATION
7-10 Sysmex SF-3000 Operator's Manual --- Revised September 1995
2.4 Exclude (Excluding and Recovering Data)
Analyzed calibration data whose compensation factor differs significantly from 100%
(due to insufficient mixing, analysis error, etc.) may be excluded from the calibration
calculation. If necessary, excluded calibration data can also be recovered.
To exclude calibration data:
(1) Use the [↑] and [↓] keys to position the cursor on the calibration data to be
excluded from the calculation.
DP:1234567890123
Sampler
Manual Mode
HGB
1
2
3
4
5
6
7
8
9
10
Reference
Analyze Compensation (%)
Avg.
Setting Analyze
Exclude Exe. Cal
Graphic Return
CALIB.
Sysmex
Not Ready
HGB HGBHCT HCT HCT
< HGB/HCT Auto >
Next No.1234567890123
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.940.0
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
[ Current ]
HGB 98.1%
HCT 100.1%
X
M
HC GP DP SU
↑
↓
Figure 7-10: Excluding Data Screen (1)
(2) From the HGB/HCT Auto submenu, press [Exclude]. The line is drawn through
the calibration data, and the cursor moves to the next sample area. "Reference",
"Analyze", and "Compensation (%)" averages are recalculated and displayed.
DP:1234567890123
Sampler
Manual Mode
HGB
1
2
3
4
5
6
7
8
9
10
Reference
Analyze Compensation (%)
Avg.
Setting Analyze
Exclude Exe. Cal
Graphic Return
CALIB.
Sysmex
Not Ready
HGB HGBHCT HCT HCT
< HGB/HCT Auto >
Next No.1234567890123
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
15.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.9
39.940.0
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.1
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
100.0
[ Current ]
HGB 98.1%
HCT 100.1%
X
M
HC GP DP SU
↑
↓
Figure 7-11: Excluding Data Screen (2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals