FUNCTIONAL DESCRIPTIONS
10-4 Sysmex SF-3000 Operator's Manual -- Revised October1995
3. SAMPLE FLOW BLOCK DIAGRAM
WBC Analysis
0.2 mL
0.967 mL
Whole Blood Aspiration
Auto Mode: 270 µL
Manual Mode: 170 µL
Sheath
Injector
Piston
0.967 mL
M
CELLPACK
STROMATOLYSER-FD (II)
Analyzed 30.4 µL of diluted
sample in WBC Detector Unit
STROMATOLYSER-FB 33.0 µL
1:30 dilution in
Reaction Chamber
STROMATOLYSER-FD (I) 33.0 µL
1:36 dilution in
Reaction Chamber
WB Pump
Sample Rotor
Valve (SRV)
Figure 10-4: WBC Sample Flow
RBC/PLT/HGB Analysis
CELLPACK
SULFOLYSER
40.0 µL
4.0 µL
Sample Rotor
Valve (SRV)
3.0 µL
Analyzed 1:500 of diluted
sample in HGB Flow Cell
Analyzed 1:25000 of diluted
sample in RBC Detector
1:500 dilution in
RBC Mixing Chamber
1.996 mL
1.96 mL
0.997 mL
0.5 mL
Figure 10-5: RBC/PLT/HGB Sample Flow