SYSTEM SETUP
Sysmex SF-3000 Operator's Manual -- Revised September 1995 11-9
2.3 Auto Erase
Use this program to start or stop the automatic erasing of the oldest data if the memory is
full when new data is stored.
To set auto erase:
(1) Press [Auto Erase] key in the Auto Management submenu.
(2) Enter the password and press [ENTER] key.
The current auto erase condition will appear on the Auto Erase setting screen. The
condition selected is backlit in light-blue.
S T O P
F I F O
< AUTO ERASE MODE >
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Quit
Output
Cond.
Output
Mode
Auto
Erase
Return
Auto
Valid.
Figure 11-7: Auto Erase Setting Window
(3) Press the appropriate key to select the condition.
The auto erase conditions are as follows:
STOP : No data is automatically erased. Data will need to be erased
manually when memory is full.
FIFO : Erase the oldest data automatically. This is the factory setting.
NOTE: The factory setting is "FIFO".