EasyManua.ls Logo

SYSMEX SF-3000 - ID BAR CODE READER ADJUSTMENT (Option)

SYSMEX SF-3000
605 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ADJUSTMENT
9-10 Sysmex SF-3000 Operator's Manual -- Revised September 1995
6. ID BAR CODE READER ADJUSTMENT (Option)
NOTE: When adjusting the ID Bar-Code Reader, contact your Sysmex
service representative for his/her instruction and assistance.
(1) Pull out the Sampler Unit by approx. 10 cm, and remove the Main Unit Front
Panel-B Assy.
CAUTION: DO NOT open the front panel or try to remove the Sampler Unit
without instruction and assistance of Sysmex service representative.
DO NOT pull the Sampler Unit more than 10 cm to avoid tubing
disconnection.
(2) Push and refit the Sampler Unit.
CAUTION: When refit the Sampler Unit, take care not to pinch the tubing.
(3) Prepare the sample racks containing blood collection tubes with the 2 mm wide ID
bar-code labels attached properly. See Figure 9-12.
Align the right end of the 2 mm-
wide barcode with the center of
the rack slit on a 15 mmOD
collection tube.
Figure 9-12: Affixing the Bar-Code Labels
(4) Perform the ID reader test program by pressing [Mainte.], [Test] and [ID Read]
keypads. When "Set Rack and press Start Switch" appears, press the Start Switch.
(Refer to Chapter 8, Section 3.7: Test Bar Code Reader for the procedures.)
(5) Verify that the ID bar-codes are read correctly.

Table of Contents

Related product manuals