CALIBRATION
7-18 Sysmex SF-3000 Operator's Manual --- Revised September 1995
4. CALIBRATION HISTORY
Up to 10 calibration histories can be accessed from the calibration history file.
4.1 Displaying the Cal. History
(1) Press [Calib.] keypad in the root menu. The Calibration screen containing
submenu will appear.
NOTE: If the [Calib.] keypad is not displayed on the LCD, press [More] to
change the root menu display.
(2) Press [Cal. History]. The message "Input Password" and numeric keypads will
appear. Input the password and press [ENTER].
(3) The history file is displayed. This file displays the dates, time, and HGB/HCT
compensation factors of the 10 latest calibrations.
6
4
5
Delete
Date
1
2
7
8
9
10
Cancel
< Cal. History >
HGB HCT
1995/08/20 10:15 99.4 101.0
3
1995/10/31 13:45 100.2 99.8
X
M DP:1234567890123
Sysmex
Ready
Sampler
Manual Mode Next No.1234567890123
HC GP DP SU
Return
Cal.
History
HGB/HCT
Manual
HGB/HCT
Auto
Figure 7-18: Calibration History Screen