MAINTENANCE & SUPPLIES, As Needed
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-33
7.3 Execute Clean RBC Transducer Sequence
If an aperture clog persists even after clog removal sequence has been executed, execute
the clean transducer sequence as described below.
WARNING! The transducer should be considered potentially biohazardous. Do
not handle the transducer without proper protective equipment.
Wear rubber gloves when cleaning. Wash your hands with an
antiseptic solution after completing the procedure.
(1) Press the [Mainte.] key from the root menu. The Maintenance second menu will be
displayed.
NOTE: If the [Mainte.] key is not displayed on the LCD, press [More] to
change the displayed root menus.
(2) Press [Service Sequence] key. The Service Sequence submenu will appear.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Clean
Waste Chamber
Clean
WBC Detector
Clean
RBC Transducer
Replace Sheath
Flow Reagent
<Service Sequence>
WBC Flow Cell
Air Bubble Removal
Return
Figure 5-7-7: Service Sequence Submenu