MAINTENANCE & SUPPLIES, As Needed, Execute Clog Removal Sequence
5-32 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(3) The message "Removing Clog" appears on the LCD.
Sysmex
Running
DP:1234567890123
Clog Removal Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Removing Clog.
Please wait.
Figure 5-7-6: Clog Removal Sequence Display
(4) After the Clog Removal sequence is completed, the instrument returns to the ready
status.
(5) Record the clog removal on the maintenance check list.
NOTE: If the error persists, execute the clean transducer sequence by
referring to the following
Section 7.3.