MAINTENANCE & SUPPLIES
5-46 Sysmex SF-3000 Operator's Manual -- Revised September 1995
8. REPLACEMENTS
8.1 Replace Reagent
When the supply of reagent runs out during an analysis, the instrument automatically
stops and one of the messages below appears. At this time replace the reagent indicated.
Table 5-8-1: Reagent Messages
MESSAGE: REPLACE:
Replace Reagent Container PK(1)
Replace Reagent Container PK(2) CELLPACK (PK-30L)
Replace Reagent Container PK(3)
Replace Reagent Container FDO STROMATOLYSER-FD (I) (FDO-200A)
Replace Reagent Container FDT STROMATOLYSER-FD (II) (FDT-200A)
Replace Reagent Container FBA STROMATOLYSER-FB (FBA-200A)
Replace Reagent Container SLS SULFOLYSER (SLS-210)
(1) Verify that the new reagent container:
is within the expiration date
does not display any signs of contamination
has been stored at room temperature.
CAUTION: Store reagents at a room temperature of 5°C to 30°C. Do not use
reagent that has been thought to be frozen.
Do not shake the reagent containers to prevent formation of micro
bubbles.
Place additional reagent cubes near the instrument prior to use to
prevent micro bubbles formation.
After breaking the seal, make sure that no dust, bacteria, or other
contaminants make their way into the container.
(2) Remove the cap from the new reagent's container. Save the cap for future use.
(3) Loosen the cap of the empty reagent container, and pull the Cubitainer spout kit
tubing straight out.
(4) Insert the spout kit tubing straight into the new reagent container, and tighten the
cap.