QUALITY CONTROL, QC Parameter Selection
6-8 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(6) Pressing [OK] key executes the new settings and the X /L-J parameter setting
window will disappear.
Pressing [Cancel] key cancels the new settings and the X /L-J parameter setting
window will disappear.
NOTE: • Parameters specified as "Not Used" will not be plotted on the QC
charts.
• Even though parameters are specified as "Not Used", the instrument
will still perform analysis on them. They are just excluded from
being plotted on the QC chart. If the operator changes "Not Used"
parameter to a "Used" parameter, all previously excluded control
data will be included on the QC chart.
4.2 Selecting QC Parameter ( X M Control)
Use this program to exclude unnecessary (non-monitored) parameters from X M QC
analysis.
(1) Press [QC] and [Settings] to select QC system setting program. The QC Settings
submenu will appear.
(2) Press [ X M Param.] keypad. The password entry screen will appear.
(3) Input the password and press [ENTER]. The X M parameters setting window
will appear as shown below.
XM Param.
OK
Cancel
WBC
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Not Used
Used
Used
Used
Used
Used
Used
Used
Used
Used
Used
Used
Used
↑
↓
←
→
Figure 6-6: XM Parameters Setting Window