QUALITY CONTROL, QC Parameter Selection
Sysmex SF-3000 Operator's Manual -- Revised September 1995 6-9
(4) Use the [↑] and [↓] keys to specify a parameter. The selected parameter is
displayed with green back-light.
(5) Use the [←] and [→] keys to specify "Used" (if the parameter is to be included in
QC analysis) or "Not Used" (if the parameter is to be excluded from QC analysis).
The selected setting is displayed in reverse video.
(6) Pressing [OK] key executes the new settings and the X
M parameter setting
window will disappear.
Pressing [Cancel] key cancels the new settings and the X M parameter setting
window will disappear.
NOTE: • Parameters specified as "Not Used" will not be plotted on the QC
charts.
• Even though parameters are specified as "Not Used", the instrument
will still perform analysis on them. They are just excluded from
being plotted on the QC chart. If the operator changes "Not Used"
parameter to a "Used" parameter, all previously excluded control
data will be included on the QC chart.