INSTALLATION
B-28 Sysmex SF-3000 Operator's Manual -- Revised September 1995
9. BATTERY BACKUP
(1) Loosen the 2 screws at the front of the right side panel. Remove the right side
panel by sliding it backward as shown in Figure B-39.
(2) Pull out the PCB No. 6358. See Figure B-40.
PCB No.6358
Figure B-39: Loosen Screws Figure B-40: Remove PCB No. 6358
(3) Connect the battery connector to the J11 of PCB No. 6358 as shown in Figure
B-41.
J11
PCB No. 6358
Battery
Figure B-41: Connect Battery Connector
(4) Install the PCB No. 6358.
(5) Install the right side panel, and secure it with two screws.