RESULT INTERPRETATION/REPORTS, Output Stored Data
4-52 Sysmex SF-3000 Operator's Manual -- Revised September 1995
11.2 Marked Data
This program specifies the marked data for output to one of the Printers or to the Host
Computer.
To output marked data:
(1) Press the [Stored Data] keypad in the Root Menu. The List Display with second
menu keys will appear.
(2) Press the [Mark] keypad to the right of List Display. Mark the data which will be
output, if necessary. Refer to Section 6: Mark Data for the procedure.
(3) Press the [More] keypad in the second menus, and press [Output] key in the altered
second menus. The Select Sample window will appear.
(4) Press [Marked] key. The Select Device window will appear.
(5) Select a device by pressing [DP], [GP Graph], [GP List] or [HC] keypad. The
output to the selected device will start.
(Press [Quit] to stop the program without outputting the data, and to return to the
List Display.)
(6) To cancel the data output after it has begun, press [Output] and then [Cancel
Output] key in the Select Sample window. Refer to Section 11.4: Cancel (Output).
NOTE: • Wait until printing is complete before specifying another output for printing.
• The marks will be cleared after the instrument power is OFF.
11.3 All Data
All the stored data can be output to one of the Printers or to the Host Computer.
To output all data:
(1) Press [All] key in the Select Sample window. The Select Device window will
appear.
(2) Select a device by pressing [DP], [GP Graph], [GP List] or [HC] keypad. The
output to the selected device will start.
(Press [Quit] to stop the program without outputting the data, and to return to the
List Display.)
(3) To cancel the data output after it has begun, press [Output] and then [Cancel
Output] key in the Select Sample window. Refer to Section 11.4: Cancel (Output).
NOTE: Wait until printing is complete before specifying another output for printing.