CALIBRATION
7-12 Sysmex SF-3000 Operator's Manual --- Revised September 1995
2.5 Exe. Cal. (Calibrating)
New compensation factors are calculated using the averages of the reference and
measured values.
NOTE: • Under [Current], the current compensation factor obtained from the
previous calibration is displayed.
• To update the [Current] compensation factor with the average found
in the "Compensation (%)" column, use the "Exe.Cal" program.
(1) From the HGB/HCT Auto submenu, press [Exe.Cal]. The message "Execute
Calibration" will appear.
< Execute Calibration >
HGB : 102.6%
HCT : 103.1%
O K
Cancel
Figure 7-14: Execute Calibration Window
Pressing [OK] key updates the "Current" column by inserting the new
compensation factor. The screen returns to the previous display.
Pressing [Cancel] key returns the instrument to the previous screen and permits the
HGB/HCT Auto program to continue without updating the calibration values.
NOTE: "Exe.Cal" can be executed only once. However, it can be executed
again after a reference value is changed, reanalysis is performed or
data is excluded.