B-26 Sysmex SF-3000 Operator's Manual -- Revised September 1995
8.4 Reagent Lines
(1) Prepare the following reagent tubes:
Tube polyurethane 4 mmID x 6 mmOD --- 5
Tube silicone 4 mmID x 8 mmOD --- 2 (for 500 mL bottles of FDT and SLS)
(2) Connect the reagent inlet nipples to the reagent containers using the polyurethane
tube (4 mmID x 6 mmOD). Use two each of silicone tube (4 mmID x 8 mmOD)
when connecting the 500 mL bottles. See the figure below.
FDO Inlet Nipple
FDT Inlet NippleFBA Inlet Nipple
PK Inlet Nipple
SLS Inlet Nipple
Silicone Tube (4x8)
STROMATOLYSER-FD(II) (FDT)
and SULFOLYSER (SLS)
Polyurethane Tube
(4x6)
Figure B-37: Reagent Connections
CAUTION: Do not spill reagent onto the instrument. If reagent does spill, wipe
out immediately with a damp cloth.
If reagent gets on clothes or skin, wash immediately with excess
water. If reagent gets into eyes, flush with water and seek medical
attention immediately.
Store reagent between of 5 - 30˚C. (Do not let reagent freeze.)
Do not let dirt, dust, or germs get into the reagent after it has been
opened. Reagent must be used within two months of opening.