INSTALLATION
B-26 Sysmex SF-3000 Operator's Manual -- Revised September 1995
8.4 Reagent Lines
(1) Prepare the following reagent tubes:
• Tube polyurethane 4 mmID x 6 mmOD --- 5
• Tube silicone 4 mmID x 8 mmOD --- 2 (for 500 mL bottles of FDT and SLS)
(2) Connect the reagent inlet nipples to the reagent containers using the polyurethane
tube (4 mmID x 6 mmOD). Use two each of silicone tube (4 mmID x 8 mmOD)
when connecting the 500 mL bottles. See the figure below.
FDO Inlet Nipple
FDT Inlet NippleFBA Inlet Nipple
PK Inlet Nipple
SLS Inlet Nipple
Silicone Tube (4x8)
STROMATOLYSER-FD(II) (FDT)
and SULFOLYSER (SLS)
Polyurethane Tube
(4x6)
Figure B-37: Reagent Connections
CAUTION: • Do not spill reagent onto the instrument. If reagent does spill, wipe
out immediately with a damp cloth.
• If reagent gets on clothes or skin, wash immediately with excess
water. If reagent gets into eyes, flush with water and seek medical
attention immediately.
• Store reagent between of 5 - 30˚C. (Do not let reagent freeze.)
• Do not let dirt, dust, or germs get into the reagent after it has been
opened. Reagent must be used within two months of opening.