SAMPLE PROCESSING, QC: Manual Closed Mode
Sysmex SF-3000 Operator's Manual -- Revised December 1995 2-17
If the error message < X Limit Error> or <L-J Limit Error> appears in the system status
area (upper portion of the LCD), the analyzed data has exceeded the control limit.
The "+" or "-" sign in yellow backlight will be displayed in the "Judgment" column and
the "Check Control Chart" message window appears. On the graphic display, the "+" or
"-" sign in yellow backlight will also be added to the parameter which is outside the
control limits. Check the control chart and review the control data. Reanalyze the
sample, if needed. Refer to Chapter 6: Quality Control.
When the message "Reanalyze Control" appears, the analyzed data is more than 3 times
the limit and will not be recorded in memory. The "+" or "-" sign in red backlight will be
displayed in the "Judgment" column. On the graphic display, the "+" or "-" sign in red
backlight will also be added to the parameter which is outside the control limits.
Reanalyze the sample.
If an analysis error occurs during QC analysis, the error message <QC Analysis Error>
will be displayed. The results are not plotted as a control data point. Reanalyze the
sample.
3.4 QC Analysis: Manual CP Mode
QC analyses can be performed in Manual Closed mode if the Manual CP Unit is installed
and SF CHECK is used. Follow the procedure below to execute quality control analyses
in Manual CP mode.
The procedures to select the Closed mode, to select QC file and the results displays are
the same as those in the Manual Closed mode. Refer to the previous Section 3.3: QC
Analysis: Manual Closed Mode for the detailed procedures.
To analyze the control material in the Manual CP mode, see Section 4.5: Manual CP
Mode in this chapter.