QUALITY CONTROL, Target/Limit, Help
6-20 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(5) The Help message screen (Page 1) will appear.
Use the [] and [] keys to change the pages. To cancel the process and exit the
program, press [Cancel] key.
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
< SF CHECK HELP (PAGE1) >
Cancel
[PRODUCT NAME]
SF CHECK; Hematology control blood
[INTENDED USE]
SF CHECK is a hematology reference control for evaluating the accuracy and
precision of the SF-3000 instrument system that measures the hematology
values. Results must be obtained through analysis in the instrument Quality
Control Analysis Mode. Do not use this material to calibrate the instrument.
[SUMMARY]
SF CHECK is to be used as a hematology control blood for the quality control
of the SF-3000 instrument system.
[PRINCIPLES]
Use of stabilized cell preparations for controlling hematology instrumentation
is an established procedure. When handled like a patient sample and assayed in
the Quality Control Mode of a properly calibrated and functioning instrument,
SF CHECK will provide values within the expected range indicated on the assay
sheet.
Figure 6-15: Help Message Screen (Page 1)
(6) Pressing the [Return] key returns the display to the screen displayed before the
[Help] key was pressed.
NOTE: SF CHECK HELP messages are subject to change without notice.