EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
MAINTENANCE & SUPPLIES, As Needed, WBC Flow Cell Air Bubble Removal
5-42 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(2) Press [Service Sequence] key. The Service Sequence submenu will appear.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Clean
Waste Chamber
Clean
WBC Detector
Clean
RBC Transducer
Replace Sheath
Flow Reagent
<Service Sequence>
WBC Flow Cell
Air Bubble Removal
Return
Figure 5-7-20: Service Sequence Submenu
(3) Press [WBC Flow Cell Air Bubble Removal] key. During the air bubbles removal
sequence, the message shown below will appear on the LCD.
Sysmex
Running
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Removing air bubbles in WBC Flow Cell.
Please wait.
Maintenance
Clean
RBC Transducer
Return
Figure 5-7-21: Removing WBC Flow Cell Air Bubbles
(4) After the rinse sequence is completed, the instrument returns to the Ready status.
(5) Keep the record on the maintenance check list.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals