MAINTENANCE & SUPPLIES, As Needed, WBC Flow Cell Air Bubble Removal
5-42 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(2) Press [Service Sequence] key. The Service Sequence submenu will appear.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Clean
Waste Chamber
Clean
WBC Detector
Clean
RBC Transducer
Replace Sheath
Flow Reagent
<Service Sequence>
WBC Flow Cell
Air Bubble Removal
Return
Figure 5-7-20: Service Sequence Submenu
(3) Press [WBC Flow Cell Air Bubble Removal] key. During the air bubbles removal
sequence, the message shown below will appear on the LCD.
Sysmex
Running
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Removing air bubbles in WBC Flow Cell.
Please wait.
Maintenance
Clean
RBC Transducer
Return
Figure 5-7-21: Removing WBC Flow Cell Air Bubbles
(4) After the rinse sequence is completed, the instrument returns to the Ready status.
(5) Keep the record on the maintenance check list.