CALIBRATION, Overview of HGB/HCT Auto Cal. program
7-4 Sysmex SF-3000 Operator's Manual --- Revised September 1995
(3) The HGB/HCT Auto screen containing submenu will appear.
Reference Analyze Compensation (%)
HGB
1
2
3
4
5
6
7
8
9
10
Avg.
HGB 98.1%
[ Current ]
Setting Analyze Exclude Exe. Cal Graphic Return
CALIB.
DP:1234567890123
Sysmex
X
M
Not Ready Manual Mode Next No.1234567890123
Sampler
< HGB / HCT Auto >
HCT HGB HCT HGB HCT
HCT 100.1%
HC GP DP SU
↓
↓
Figure 7-3: HGB/HCT Auto Screen
NOTE: When the HGB/HCT Auto program is in use, the Main Unit READY
LED turns off. Normal samples cannot be analyzed at this time.
Submenu Keypads Operation
[Setting] : Enters the reference HGB and HCT values. Refer to the following
Section 2.2 Setting.
[Analyze] : Analyzes the samples. Refer to Section 2.3 Analyze.
[Exclude] : Review data obtained after specimens have been analyzed with this
program. Data may be marked and not included in the final calibration
calculation. Data selected for exclusion may also be recovered before
auto calibration is executed. Refer to Section 2.4 Exclude
[Exe.Cal] : Calculates the changes necessary to update the [Current] compensation
factors. Data from the difference between reference values and
analyzed samples is utilized. Refer to Section 2.5 Exe.Cal.
[Graphic] : Displays the scattergrams and particle size distributions of the last
sample analyzed. Refer to Section 2.6 Graphic.