MAINTENANCE & SUPPLIES, Monthly, Clean Waste Chamber
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-17
(2) Press [Service Sequence] key. The Service Sequence submenu will appear.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Clean
Waste Chamber
Clean
WBC Detector
Clean
RBC Transducer
Replace Sheath
Flow Reagent
<Service Sequence>
WBC Flow Cell
Air Bubble Removal
Return
Figure 5-5-5: Service Sequence Submenu
(3) Press [Clean Waste Chamber] key. The message "It will take approx. 20 min.
Press Start Switch." appears.
Pressing [Cancel] cancels the program without cleaning and returns to the Service
Sequence submenu display.
Sysmex
Ready
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
< Clean Waste Chamber >
It will take approx. 20 min.
Press Start Switch.
Cancel
Maintenance
Return
Figure 5-5-6: Clean Waste Chamber Execute Message