EasyManua.ls Logo

SYSMEX SF-3000 - Page 189

SYSMEX SF-3000
605 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MAINTENANCE & SUPPLIES, Monthly, Clean Waste Chamber
5-18 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(4) Press the Start Switch. The reagent is discharged from the WBC Reaction
Chambers.
Sysmex
Running
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Draining reagent in the Reaction Chambers.
Please wait.
Maintenance
Clean
RBC Transducer
Return
Figure 5-5-7: Cleaning Waste Chamber Message (1)
(5) When the reagent is discharged from the Reaction Chambers, the message "Place 1
mL each of CELLCLEAN both in the Reaction Chambers 1 and 2. Press Start
Switch." will appear.
Sysmex
Ready
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Maintenance
<Clean Waste Chamber>
Place 1 mL each of CELLCLEAN both
in the Reaction Chambers 1 and 2.
Press Start Switch.
Return
Figure 5-5-8: Cleaning Waste Chamber Message (2)

Table of Contents

Related product manuals