MAINTENANCE & SUPPLIES, Monthly, Clean Waste Chamber
5-18 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(4) Press the Start Switch. The reagent is discharged from the WBC Reaction
Chambers.
Sysmex
Running
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Draining reagent in the Reaction Chambers.
Please wait.
Maintenance
Clean
RBC Transducer
Return
Figure 5-5-7: Cleaning Waste Chamber Message (1)
(5) When the reagent is discharged from the Reaction Chambers, the message "Place 1
mL each of CELLCLEAN both in the Reaction Chambers 1 and 2. Press Start
Switch." will appear.
Sysmex
Ready
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Maintenance
<Clean Waste Chamber>
Place 1 mL each of CELLCLEAN both
in the Reaction Chambers 1 and 2.
Press Start Switch.
Return
Figure 5-5-8: Cleaning Waste Chamber Message (2)