MAINTENANCE & SUPPLIES, Monthly, Clean Orifice
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-15
(4) Rinse the orifice with tap water.
Figure 5-5-3: Rinse Orifice
(5) Check the orifice for dirt, and wipe off any water.
(6) Reattach the orifice to its original position.
(7) Reset the counter of the Vacuum Orifice. Refer to Section 9: Reset the Counter in
this chapter.
NOTE: Make sure to reset the counter after cleaning. If not, the message
"Clean the Orifice" will remain on the LCD.
(8) Record the cleaning on the maintenance check list.