76 Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
Key Entry Atten Hold Off On
Remarks Refer to the “:ALC:LEVel” on page 66.
:ATTenuation:BYPass
Supported All Models
[:SOURce]:POWer:ATTenuation:BYPass ON|OFF|1|0
[:SOURce]:POWer:ATTenuation:BYPass?
This command enables or disables the attenuator bypass setting. The attenuator hold mode must be
enabled to use this command.
ON (1) This selection allows the signal generator’s automatic level control (ALC) to adjust
the attenuator hold mode. Output power is controlled solely by the ALC setting.
OFF (0) This choice allows for a user–selected attenuator setting combined with the ALC
setting.
*RST 0
Key Entry Atten Bypass Off On
:HARMonics
Supported All Models
[:SOURce]:POWer:HARMonics[:STATe] ON|OFF|1|0
[:SOURce]:POWer:HARMonics?
This command enables or disables the optimize harmonics setting. The optimize harmonics mode
modifies the attenuator and automatic level control (ALC) settings to give optimal harmonics
performance. Optimize harmonics mode does not change the RF output power. The attenuator hold
mode cannot be enabled while this mode is active, and modulations cannot be enabled while this
mode is active.
ON (1) This selection allows the signal generator to optimize harmonics by modifying the
attenuator and automatic level control settings.
OFF (0) This selection disables the optimize harmonics mode.
*RST 0
Key Entry Optimize Harmonics
[:LEVel][:IMMediate]:OFFSet
Supported All Models
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet <value><unit>
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet?
This command sets the power offset value.
*RST 0.00000000E000
Range 200 to 200 dB