48 Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Frequency Subsystem ([:SOURce])
*RST +1.0000000000000E+07 Hz
Range +1.0000000000000E+06 to +5.0000000000000E+07 Hz
Key Entry Ref Oscillator Ext Freq
Remarks If the entered frequency does not match the frequency of the entered reference,
an unlocked condition will occur and an error message will appear.
:ROSCillator:OVEN:STATe
Supported All Models
[:SOURce]:ROSCillator:OVEN:STATe ON|OFF|1|0
[:SOURce]:ROSCillator:OVEN:STATe?
This command turns the 10 MHz oven oscillator on or off.
*RST 1
Key Entry Oven Oscillator On Off
:ROSCillator:OVEN:TUNE
Supported All Models
[:SOURce]:ROSCillator:OVEN:TUNE <value>
[:SOURce]:ROSCillator:OVEN:TUNE?
This command tunes the internal oven oscillator frequency.
The user value offsets the factory tuned value (the value is added to the factory calibrated DAC
value). The tune value of 0 sets the factory calibrated value.
*RST +0.00000000E+000
Range ï€4096 to +4096
Key Entry Oven Oscillator Tune
:ROSCillator:SOURce
Supported All Models
[:SOURce]:ROSCillator:SOURce INTernal|EXTernal|BBG
[:SOURce]:ROSCillator:SOURce?
This command sets the current reference oscillator source: INT (internal), EXT (external), or BBG
(internal baseband generator).
:ROSCillator:SOURce:AUTO
Supported All Models
[:SOURce]:ROSCillator:SOURce:AUTO ON|OFF|1|0
[:SOURce]:ROSCillator:SOURce:AUTO?
This command enables or disables the ability of the signal generator to automatically select between
the internal and an external reference oscillator.