Agilent X-Series Signal Generators SCPI Command Reference 157
System Commands
Status Subsystem (:STATus)
:QUEStionable:ENABle
Supported All Models
:STATus:QUEStionable:ENABle <value>
:STATus:QUEStionable:ENABle?
This command determines which bits in the Data Questionable Event Register will set the Data
Questionable Status Group Summary bit (bit 3) in the Status Byte Register.
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:FREQuency:CONDition?
Supported All Models
:STATus:QUEStionable:FREQuency:CONDition?
This query returns the decimal sum of the bits in the Data Questionable Frequency Condition
Register. For example, if the 1 GHz internal reference clock is unlocked (bit 2), a value of 4 is
returned.
Range 0 to 32767
Remarks The data in this register is continuously updated and reflects current conditions.
Refer to the Programming Guide for more information.
:QUEStionable:FREQuency:ENABle
Supported All Models
:STATus:QUEStionable:FREQuency:ENABle <value>
:STATus:QUEStionable:FREQuency:ENABle?
This command determines which bits in the Data Questionable Frequency Event Register will set the
frequency summary bit (bit 5) in the Data Questionable Condition Register.
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.