Agilent X-Series Signal Generators SCPI Command Reference 115
System Commands
IEEE 488.2 Common Commands
*SAV
Supported All Models
*SAV <reg>,<seq>
The Save (SAV) command saves signal generator settings to the specified memory register <reg> of
the specified sequence <seq>.
Remarks The save function does not save all signal generator settings. Refer to the User’s
Guide for more information on the save function. Refer to “*RCL” on page 114 for
information on recalling saved signal generator settings.
*SRE
Supported All Models
*SRE <data>
The Service Request Enable (SRE) command sets the value of the Service Request Enable Register.
The variable <data> is the decimal sum of the bits that will be enabled. Bit 6 (value 64) is ignored
and cannot be set by this command.
Range 0 to 255
Remarks Refer to the Programming Guide for more information.
Entering values from 64 to 127 is equivalent to entering values from 0 to 63.
The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset it to zero.
*SRE?
Supported All Models
*SRE?
The Service Request Enable (SRE) query returns the value of the Service Request Enable Register.
Range 0 to 63 or 128 to 191
Remarks Refer to the Programming Guide for more information.
Range Registers: 0 to 99 Sequences: 0 to 9
Key Entry Save Reg Save Seq[n] Reg[nn]

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals