EasyManua.ls Logo

Agilent Technologies N5171B

Agilent Technologies N5171B
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Agilent X-Series Signal Generators SCPI Command Reference 115
System Commands
IEEE 488.2 Common Commands
*SAV
Supported All Models
*SAV <reg>,<seq>
The Save (SAV) command saves signal generator settings to the specified memory register <reg> of
the specified sequence <seq>.
Remarks The save function does not save all signal generator settings. Refer to the User’s
Guide for more information on the save function. Refer to “*RCL” on page 114 for
information on recalling saved signal generator settings.
*SRE
Supported All Models
*SRE <data>
The Service Request Enable (SRE) command sets the value of the Service Request Enable Register.
The variable <data> is the decimal sum of the bits that will be enabled. Bit 6 (value 64) is ignored
and cannot be set by this command.
Range 0 to 255
Remarks Refer to the Programming Guide for more information.
Entering values from 64 to 127 is equivalent to entering values from 0 to 63.
The setting enabled by this command is not affected by signal generator preset or
*RST. However, cycling the signal generator power will reset it to zero.
*SRE?
Supported All Models
*SRE?
The Service Request Enable (SRE) query returns the value of the Service Request Enable Register.
Range 0 to 63 or 128 to 191
Remarks Refer to the Programming Guide for more information.
Range Registers: 0 to 99 Sequences: 0 to 9
Key Entry Save Reg Save Seq[n] Reg[nn]

Table of Contents

Related product manuals