Agilent X-Series Signal Generators SCPI Command Reference 21
Basic Function Commands
Correction Subsystem ([:SOURce]:CORRection)
:FLATness:PAIR
Supported All Models
[:SOURce]:CORRection:FLATness:PAIR <freq.>[<freq suffix>],
<corr.>[<corr suffix>]
This command sets a frequency and amplitude correction pair.
The maximum number of points that can be entered is 3,201.
<corr.> This variable is the power correction.
Range Frequency range varies and is model dependent. Refer to the instrument’s Data
Sheet.
Key Entry Configure Cal Array
:FLATness:POINts
Supported All Models
[:SOURce]:CORRection:FLATness:POINts?
This query returns the number of points in the user–flatness correction file.
:FLATness:PRESet
Supported All Models
CAUTION The current correction data will be overwritten once this command is executed. Save the
current data if needed. Refer to :FLATness:STORe command for storing user–flatness
files.
[:SOURce]:CORRection:FLATness:PRESet
This command presets the user–flatness correction to a factory–defined setting that consists of one
point.
Key Entry Preset List
:FLATness:STEP:POINts
Supported All Models
[:SOURce]:CORRection:FLATness:STEP:POINts <points> |MAXimum|MINimum|DEFault|
[:SOURce]:CORRection:FLATness:STEP:POINts?[MAXimum|MINimum]
This command is used to define the number of points in the user flatness calibration step array.
See also :FLATness:STEP:STARt and :FLATness:STEP:STOP commands.
*RST 2
Range 2 to 3,201
Key Entry # Points