Agilent X-Series Signal Generators SCPI Command Reference 221
Analog Modulation Commands
Low Frequency Output Subsystem ([:SOURce]:LFOutput)
This command sets the noise type when :SOURce is set to NOISe[1] or NOISe2.
*RST POS
:OFFset
Supported All Models with Option UNT
[:SOURce]:LFOutput:OFFset <value><unit>
[:SOURce]:LFOutput:OFFset?
This command sets the DC offset (in volts) of the signal at the LF Output connector.
*RST 0.000 V
:SOURce
Supported All Models with Options UNT and 303
[:SOURce]:LFOutput:SOURce MONitor|FUNCtion[1]|FUNCtion2|SWEep|DUAL|NOISe[1]|NOISe2|DC
[:SOURce]:LFOutput:SOURce?
This command selects the source for the LF output.
MONitor Selects monitoring on the LF output BNC. Select the monitoring source using the
:SOURce:MONitor command.
FUNCtion[1] Selects function generator 1 as the modulation source.
FUNCtion[2] Selects function generator 2 as the modulation source.
SWEep Selects the swept function generator as the LF output BNC source. If AM or FM or
PM is modulating the swept function generator then the LF output BNC will have
the unmodulated signal if you choose to monitor the swept function generator.
DUAL Selects the dual function generator as the modulation source.
NOISe[1] Selects noise generator 1 as the modulation source.
NOISe2 Selects noise generator 2 as the modulation source.
DC Selects a DC voltage level as the LF output BNC source.
*RST MONitor
Remarks Only MON, FUNC1 and DC are available without Option 303.
:SOURce:MONitor
Supported All models with Option UNT
[:SOURce]:LFOutput:SOURce:MONitor FUNCtion[1]|FUNCtion2|SWEep|DUAL
Key Entry Uniform Gaussian
Range - 10.000 to 10.000 V
Key Entry LF Out DC Offset Into 50 Ohms LF Out DC Offset Into 1 MOhms
Key Entry Int Monitor Func Gen 1 Dual Func Gen Noise Gen 1
DC Func Gen 2 Swept Func Gen Noise Gen 2