144 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Output Subsystem (:OUTPut)
:MODulation[:STATe]
Supported All Models
:OUTPut:MODulation[:STATe] ON|OFF|1|0
:OUTPut:MODulation[:STATe]?
This command enables or disables the modulation of the RF output with the currently active
modulation type(s).
*RST 1
Key Entry Mod On/Off
Remarks Some modulation types can be simultaneously enabled such as pulse and AM.
An annunciator on the signal generator is always displayed to indicate whether
modulation is switched on or off.
[:STATe]
Supported All Models
:OUTPut[:STATe] ON|OFF|1|0
:OUTPut[:STATe]?
This command enables or disables the RF output.
*RST 0
Key Entry RF On/Off
Remarks Although you can configure and engage various modulations, no signal is available
at the RF OUTPUT connector until this command is executed.
An annunciator is always displayed on the signal generator to indicate whether
the RF output is switched on or off.