EasyManuals Logo

Agilent Technologies N5171B User Manual

Agilent Technologies N5171B
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #174 background imageLoading...
Page #174 background image
144 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Output Subsystem (:OUTPut)
:MODulation[:STATe]
Supported All Models
:OUTPut:MODulation[:STATe] ON|OFF|1|0
:OUTPut:MODulation[:STATe]?
This command enables or disables the modulation of the RF output with the currently active
modulation type(s).
*RST 1
Key Entry Mod On/Off
Remarks Some modulation types can be simultaneously enabled such as pulse and AM.
An annunciator on the signal generator is always displayed to indicate whether
modulation is switched on or off.
[:STATe]
Supported All Models
:OUTPut[:STATe] ON|OFF|1|0
:OUTPut[:STATe]?
This command enables or disables the RF output.
*RST 0
Key Entry RF On/Off
Remarks Although you can configure and engage various modulations, no signal is available
at the RF OUTPUT connector until this command is executed.
An annunciator is always displayed on the signal generator to indicate whether
the RF output is switched on or off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals