200 Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Amplitude Modulation Subsystem–Option UNT ([:SOURce])
*RST POS
:AM[1]|2:INTernal:NOISe[1]|2:TYPe
Supported All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:NOISe[1]|2:TYPe UNIForm|GAUSsian
[:SOURce]:AM[1]|2:INTernal:NOISe[1]|2:TYPe?
This command sets the noise type when :AM[1]|2:SOURce is set to NOISe[1] or NOISe2.
*RST POS
:AM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STARt
Supported All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STARt <val><units>
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STARt?
This command sets the start frequency for the swept function generator.
*RST 400.0 Hz
Range 100 mHz to 6.250 MHz
Key Entry AM Start Freq
:AM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STOP
Supported All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STOP <val><units>
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:FREQuency:STOP?
This command sets the stop frequency for the swept function generator.
*RST 400.0 Hz
Range 100 mHz to 6.250 MHz
Key Entry AM Stop Freq
:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe
Supported All Models with Options UNT and 303
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe SINE|TRIangle|SQUare|RAMP
[:SOURce]:AM[1]|2:INTernal:SWEep:FUNCtion:SHAPe?
This command sets the AM waveform type for the swept function generator.
*RST SINE
Key Entry Pos Ramp Neg Ramp
Key Entry Noise Gen 1 Noise Gen 2
Key Entry AM Sweep Waveform