Agilent X-Series Signal Generators SCPI Command Reference 53
Basic Function Commands
List/Sweep Subsystem ([:SOURce])
Entering a value with this command will have no effect, unless MANual is the selected mode. Refer
to the :LIST:MODE command for setting the proper mode.
If the point selected is beyond the length of the longest enabled list, then the point will be set to the
maximum possible point, and an error will be generated.
Range List Sweep: 1 to 3,201 Step Sweep: 2 to 65535
Key Entry Manual Point
:LIST:MODE
Supported All Models
[:SOURce]:LIST:MODE AUTO|MANual
[:SOURce]:LIST:MODE?
This command sets the operating mode for the current list or step sweep.
AUTO This choice enables the selected sweep type to perform a sweep of all points.
MANual This choice enables you to select a single sweep point. The selected point controls
the frequency and/or amplitude according to the sweep type. Refer to the
:LIST:MANual command for selecting a sweep point.
*RST AUTO
Key Entry Manual Mode Off On
:LIST:OPTions
Supported All Models
[:SOURce]:LIST:OPTions <val>,<val>
[:SOURce]:LIST:OPTions?
[:SOURce]:LIST:OPTions:POINts?
This command enables specific options during a list sweep. The command adds the capability to
suppress FM, M, and AM on any list sweep point. Additionally, frequency, power, and/or a waveform
transition can be suppressed resulting in no synthesizer, no output, or no waveform playing
interruption during a transition.