214 Agilent X-Series Signal Generators SCPI Command Reference
Analog Modulation Commands
Frequency Modulation Subsystem–Option UNT ([:SOURce])
:FM[1]|2:INTernal:SWEep:TRIGger:INTernal:SOURce
Supported All Models with Options UNT and 303
[:SOURce]:FM[1]|2:INTernal:SWEep:TRIGger:INTernal:SOURce PVIDeo|PSYNc
[:SOURce]:FM[1]|2:INTernal:SWEep:TRIGger:INTernal:SOURce?
This command selects the internal trigger source for the FM swept- sine waveform.
PVIDeo This choice selects Pulse Video as the internal trigger source for triggering sweep,
point and function generator sweeps.
PSYNc This choice selects Pulse Sync as the internal trigger source for triggering sweep,
point and function generator sweeps.
*RST PSYN
:FM[1]|2:INTernal:SWEep:TRIGger:SLOPe
Supported All Models with Options UNT and 303
[:SOURce]:FM[1]|2:INTernal:SWEep:TRIGger:SLOPe POSitive|NEGative
[:SOURce]:FM[1]|2:INTernal:SWEep:TRIGger:SLOPe?
This command sets the polarity of an external signal at the TRIG 1, TRIG 2, or PULSE BNC (see
:FM[1]|2:INTernal:SWEep:TRIGger:EXTernal:SOURce) or internal Pulse Video or Pulse Sync signal (see
:FM[1]|2:INTernal:SWEep:TRIGger:INTernal:SOURce) that will trigger a list or step sweep.
POSitive The signal generator triggers an event when it detects a rising edge on the source
signal
NEGative The signal generator triggers an event when it detects a falling edge on the source
signal
*RST POS
Key Entry Int/Ext Trigger Polarity Neg Pos
:FM[1]|2:SOURce
Supported All Models with Option UNT
[:SOURce]:FM[1]|2:SOURce FUNCtion[1]|FUNCtion2|SWEep|DUAL|NOISe[1]|NOISe2|EXT[1]|EXT2
[:SOURce]:FM[1]|2:SOURce?
This command sets the source to generate the frequency modulation.
FUNCtion[1] Selects function generator 1 as the modulation source.
FUNCtion[2] Selects function generator 2 as the modulation source.
SWEep Selects the swept function generator as the modulation source.
DUAL Selects the dual function generator as the modulation source.
NOISe[1] Selects noise generator 1 as the modulation source.
Key Entry Pulse Video Pulse Sync