Agilent X-Series Signal Generators SCPI Command Reference 205
Analog Modulation Commands
Amplitude Modulation Subsystem–Option UNT ([:SOURce])
The externally applied, ac- coupled input signal is tested for a voltage level and a
display annunciator will report a high or low condition if that voltage is > 3% of
1 V
p
.
Only FUNC1, FUNC2, EXT1 and EXT2 are available without Option 303.
:AM[1]|2:STATe
Supported All Models with Option UNT
[:SOURce]:AM[1]|2:STATe ON|OFF|1|0
[:SOURce]:AM[1]|2:STATe?
This command enables or disables the amplitude modulation for the selected path.
*RST 0
Key Entry AM Off On
Remarks The RF carrier is modulated when you have set the signal generator’s modulation
state to ON, see :MODulation[:STATe] command for more information.
Whenever amplitude modulation is enabled, the AM annunciator is turned on in
the display.
:AM[1]|2:TYPE
Supported All models with Option UNT
[:SOURce]:AM[1]|2:TYPE LINear|EXPonential
[:SOURce]:AM[1]|2:TYPE?
This command sets the AM type to linear or exponential AM.
LINear This choice selects linear AM type with depth values in units of percent/volt.
EXPonential This choice selects exponential AM type with depth values in units of dB/volt.
Example
:AM2:TYPE EXP
The preceding example selects exponential type depth values for AM path 2.
*RST LIN
Key Entry AM Type LIN EXP
:AM:WIDeband:STATe
Supported All Models with Option UNT
[:SOURce]:AM:WIDeband:STATe ON|OFF|1|0
[:SOURce]:AM:WIDeband:STATe?