Agilent X-Series Signal Generators SCPI Command Reference 159
System Commands
Status Subsystem (:STATus)
:QUEStionable:MODulation:CONDition?
Supported All Models
:STATus:QUEStionable:MODulation:CONDition?
This command returns the decimal sum of the bits in the Data Questionable Modulation Condition
Register. For example, if the modulation is uncalibrated (bit 4), a value of 16 is returned.
Range 0 to 32767
Remarks The data in this register is continuously updated and reflects current conditions.
Refer to the Programming Guide for more information.
:QUEStionable:MODulation:ENABle
Supported All Models
:STATus:QUEStionable:MODulation:ENABle <val>
:STATus:QUEStionable:MODulation:ENABle?
This command determines which bits in the Data Questionable Modulation Event Register will set the
modulation summary bit (bit 7) in the Data Questionable Condition Register.
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:MODulation:NTRansition
Supported All Models
:STATus:QUEStionable:MODulation:NTRansition <val>
:STATus:QUEStionable:MODulation:NTRansition?
This command determines which bits in the Data Questionable Modulation Condition Register will set
the corresponding bit in the Data Questionable Modulation Event Register when that bit has a
negative transition (1 to 0).
The variable <val> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.