Agilent X-Series Signal Generators SCPI Command Reference 99
System Commands
Calibration Subsystem (:CALibration)
:IQ:[:USER]
Supported N5172B with Option 653 or 655, N5182B with Option 656 or 657
:CALibration:IQ[:USER]
This command performs a IQ calibration according to the IQ calibration type. For information on
selecting the type of IQ calibration, refer to “:IQ:TYPE” on page 98.
This calibration should be run when the ambient temperature has varied by at least ±5 degrees
celsius from the ambient temperature at which the previous calibration was run.
Key Entry Execute Cal