Agilent X-Series Signal Generators SCPI Command Reference 81
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
:REFerence:STATe
Supported All Models
[:SOURce]:POWer:REFerence:STATe ON|OFF|1|0
[:SOURce]:POWer:REFerence:STATe?
This command enables or disables the RF output reference.
Once the reference state is ON, all subsequent output power settings are set relative to the reference
value.
ON (1) This choice will set the power reference state to ON. The unit displayed for
commands, :ANNotation:AMPLitude:UNIT and :POWer will be expressed in dB.
OFF (0) This choice will set the power reference state to OFF.
*RST 0
Key Entry Amptd Ref Off On
Remarks Amplitude offsets can be used with the amplitude reference mode.
:STARt
Supported All Models
[:SOURce]:POWer:STARt <value><unit>
[:SOURce]:POWer:STARt?
This command sets the first amplitude point in a step sweep.
*RST ï€1.10000000E002 (Standard) and ï€1.44000000E002 (Option 1EQ)
Range Refer to the [:LEVel][:IMMediate][:AMPLitude] command for the output power
ranges.
Key Entry Amptd Start
:STOP
Supported All Models
[:SOURce]:POWer:STOP <value><unit>
[:SOURce]:POWer:STOP?
This command sets the last amplitude point in a step sweep.
*RST ï€1.10000000E002 (Standard) and ï€1.44000000E002 (Option 1EQ)
Range Refer to the [:LEVel][:IMMediate][:AMPLitude] command for the output power
ranges.
Key Entry Amptd Stop