188 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Trigger Subsystem
Trigge r Subsystem
:ABORt
Supported All
:ABORt
This command causes the List or Step sweep in progress to abort. If INIT:CONT[:ALL] is set to on,
the sweep will immediately re–initiate. The pending operation flag affecting *OPC, *OPC?, and *WAI
will undergo a transition once the sweep has been reset.
:INITiate:CONTinuous[:ALL]
Supported All
:INITiate:CONTinuous[:ALL] ON|OFF|1|0
:INITiate:CONTinuous[:ALL]?
This command selects either a continuous or single list or step sweep. Execution of this command
does not affect a sweep in progress.
ON (1) This choice selects continuous sweep where, after the completion of the previous
sweep, the current sweep will restart automatically or wait until the appropriate
trigger source is received.
OFF (0) This choice selects a single sweep. Refer to :INITiate[:IMMediate][:ALL] for single
sweep triggering information.
*RST 0
Key Entry Sweep Repeat Single Cont
Remarks Execution of this command will not affect a sweep in progress.
:INITiate[:IMMediate][:ALL]
Supported All
:INITiate[:IMMediate][:ALL]
This command either sets or sets and starts a single List or Step sweep, depending on the trigger
type. The command performs the following:
• arms a single sweep when BUS, EXTernal, or KEY is the trigger source selection
• arms and starts a single sweep when IMMediate is the trigger source selection
This command is ignored if a sweep is in progress. See :INITiate:CONTinuous[:ALL] command for
setting continuous or single sweep. See :TRIGger[:SEQuence]:SOURce command to select the trigger
source.
Key Entry Single Sweep