EasyManuals Logo

Agilent Technologies N5171B User Manual

Agilent Technologies N5171B
458 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #96 background imageLoading...
Page #96 background image
66 Agilent X-Series Signal Generators SCPI Command Reference
Basic Function Commands
Power Subsystem ([:SOURce]:POWer)
Power Subsystem ([:SOURce]:POWer)
:ALC:BANDwidth
Supported All Models
[:SOURce]:POWer:ALC:BANDwidth|BWIDth <num>freq suffix]
[:SOURce]:POWer:ALC:BANDwidth|BWIDth?
This command overrides the signal generator's automatic ALC bandwidth selection with the users
specific selection. For waveforms with varying amplitudes, high crest factors, or both, the
recommended ALC loop bandwidth is the low bandwidth setting of the generator. Limiting the loop
bandwidth of the ALC circuit will prevent the ALC from sampling the fast rising edges of pulsed
waveforms. A limited, or narrow bandwidth will result in a longer ALC sample time and a more
accurate representation of the signal's level.
*RST 200 (2kHz when ALC:BAND:BWID is set to AUTO)
Remarks Use this command when the ALC state is set to On. This command will override
the automatic ALC bandwidth selection set by the :ALC:BANDwidth|BWIDth:AUTO
command.
:ALC:BANDwidth|BWIDth:AUTO
Supported All Models
[:SOURce]:POWer:ALC:BANDwidth|BWIDth:AUTO ON|OFF|1|0
[:SOURce]:POWer:ALC:BANDwidth|BWIDth:AUTO?
This command turns the bandwidth (BW) auto state on or off.
The bandwidth auto function allows the signal generator to automatically select a bandwidth for the
automatic leveling control (ALC) circuit.
ON (1) This choice allows the signal generator to automatically select an ALC BW. The
selection of the ALC BW depends on the signal generator modulation type.
OFF (0) This choice disables automatic selection of the ALC BW.
*RST 1
Key Entry Auto
Remarks For more information on ALC bandwidth, refer to the User’s Guide.
:ALC:LEVel
Supported All Models
[:SOURce]:POWer:ALC:LEVel <value><unit>
[:SOURce]:POWer:ALC:LEVel?
This command sets the automatic leveling control (ALC) level. Use this command after setting the
attenuation auto mode to On. Refer to :ATTenuation:AUTO command for setting the attenuation auto
mode.
Key Entry Auto 200 Hz 2 kHz 20 kHz

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies N5171B and is the answer not in the manual?

Agilent Technologies N5171B Specifications

General IconGeneral
BrandAgilent Technologies
ModelN5171B
CategoryComputer Hardware
LanguageEnglish

Related product manuals