Agilent X-Series Signal Generators SCPI Command Reference 391
Real–Time Commands
Custom Subsystem–Option 431([:SOURce]:RADio:CUSTom)
Remarks To change the modulation type, refer to “:MODulation[:TYPE]” on page 392.
Refer to “:SRATe” on page 400 for a list of the minimum and maximum symbol
rate values.
To set an asymmetric FSK deviation value, refer to the User’s Guide for more
information.
:MODulation:MSK[:PHASe]
Supported N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:MODulation:MSK[:PHASe] <val>
[:SOURce]:RADio:CUSTom:MODulation:MSK[:PHASe]?
This command sets the MSK phase deviation value.
The variable <val> is expressed in units of degrees.
*RST
+9.00000000E+001
Range 0–100
Key Entry Phase Dev
:MODulation:UFSK
Supported N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:MODulation:UFSK "<file name>"
[:SOURce]:RADio:CUSTom:MODulation:UFSK?
This command selects a user- defined FSK file from the signal generator memory.
Key Entry User FSK
Remarks
The user-defined FSK file is held in signal generator memory until the command that
selects user FSK as the modulation type is sent. Refer to “:MODulation[:TYPE]” on
page 392 to change the current modulation type.
Refer to “File Name Variables” on page 13 for information on the file name syntax.
:MODulation:UIQ
Supported N5172B/82B with Option 431
[:SOURce]:RADio:CUSTom:MODulation:UIQ "<file name>"
[:SOURce]:RADio:CUSTom:MODulation:UIQ?
This command selects a user- defined I/Q file from the signal generator memory.
Key Entry User I/Q
Remarks
The user-defined I/Q file is held in signal generator memory until the command that
selects user I/Q as the modulation type is sent. Refer to “:MODulation[:TYPE]” on
page 392 to change the current modulation type.