Agilent X-Series Signal Generators SCPI Command Reference 45
Basic Function Commands
Frequency Subsystem ([:SOURce])
[:SOURce]:FREQuency:OFFSet:STATe ON|OFF|1|0
[:SOURce]:FREQuency:OFFSet:STATe?
This command enables or disables the offset frequency.
*RST 0
Key Entry Freq Offset
Remarks Entering OFF (0) will set the frequency offset to 0 Hz.
:FREQuency:REFerence
Supported All Models
[:SOURce]:FREQuency:REFerence <value><unit>
[:SOURce]:FREQuency:REFerence?
This command sets the output reference frequency.
*RST 0.0000000000000E00
Range The range is model/option dependent. Refer to the instrument’s Data Sheet.
Key Entry Freq Ref Set
:FREQuency:REFerence:SET
Supported All Models
[:SOURce]:FREQuency:REFerence:Set
This command sets the current CW output frequency, along with any offset, as a 0 hertz reference
value.
*RST 0.0000000000000E00
Key Entry Freq Ref Set
:FREQuency:REFerence:STATe
Supported All Models
[:SOURce]:FREQuency:REFerence:STATe ON|OFF|1|0
[:SOURce]:FREQuency:REFerence:STATe?
This command enables or disables the frequency reference mode.
When the frequency reference mode is on, subsequent frequency parameters are set relative to the
reference value.
*RST 0
Key Entry Freq Ref Off On
:FREQuency:SPAN
Supported All Models
[:SOURce]:FREQuency:SPAN <num>[<freq_suffix>]|UP|DOWN