136 Agilent X-Series Signal Generators SCPI Command Reference
System Commands
Memory Subsystem (:MEMory)
:EXPort[:ASCii]:SEParator:DECimal
Supported All with Option 320 and either UNU or UNW
:MEMory:EXPort[:ASCii]:SEParator:DECimal DOT|COMMa
:MEMory:EXPort[:ASCii]:SEParator:DECimal?
This command selects whether the decimal point is a "." or a "," during export of CSV/ASCII files.
This value is persistent across preset/recall and power cycles. (At the factory the MXG is set to DOT
(“.”).) Refer to :EXPort[:ASCii]:SEParator:COLumn.
Key Entry Export Decimal Separator
:FREE[:ALL]
Supported All Models
:MEMory:FREE[:ALL]?
This command returns the number of bytes left in the non–volatile user file system.
Key Entry All
:IMPort[:ASCii]:PTRain
Supported All with Option 320 and either UNU or UNW
:MEMory:IMPort[:ASCii]:PTRain <"filename">
This command reads a CSV/ASCII file from the BINARY directory. The user must specify any
extender (such as .csv or .txt) used when placing the file into the instrument. Note that the form of
these files must be On Time<column separator>Off Time<column
separator>Repetitions<newline> or On Time<column separator>Off Time<newline> with
repetition count assumed to always be 1 in the second case. Refer to
:IMPort[:ASCii]:SEParator:DECimal.
Key Entry Import From Selected File
:IMPort[:ASCii]:SEParator:DECimal
Supported All with Option 320 and either UNU or UNW
:MEMory:IMPort[:ASCii]:SEParator:DECimal DOT|COMMa
:MEMory:IMPort[:ASCii]:SEParator:DECimal?
This command selects whether the decimal point is a "." or a "," during import of CSV/ASCII files.
This value is persistent across preset/recall and power cycles. (At the factory the MXG is set to DOT
(“.”).) Refer to :IMPort[:ASCii]:PTRain.
Key Entry Import Decimal Separator