Agilent X-Series Signal Generators SCPI Command Reference 161
System Commands
Status Subsystem (:STATus)
:QUEStionable:POWer:CONDition?
Supported All Models
:STATus:QUEStionable:POWer:CONDition?
This query returns the decimal sum of the bits in the Data Questionable Power Condition Register.
For example, if the RF output signal is unleveled (bit 1), a value of 2 is returned.
Range 0 to 32767
Remarks The data in this register is continuously updated and reflects current conditions.
Refer to the Programming Guide for more information.
:QUEStionable:POWer:ENABle
Supported All Models
:STATus:QUEStionable:POWer:ENABle <value>
:STATus:QUEStionable:POWer:ENABle?
This command determines which bits in the Data Questionable Power Event Register will set the
power summary bit (bit 3) in the Data Questionable Condition Register.
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.
:QUEStionable:POWer:NTRansition
Supported All Models
:STATus:QUEStionable:POWer:NTRansition <value>
:STATus:QUEStionable:POWer:NTRansition?
This command determines which bits in the Data Questionable Power Condition Register will set the
corresponding bit in the Data Questionable Power Event Register when that bit has a negative
transition (1 to 0).
The variable <value> is the sum of the decimal values of the bits that you want to enable.
Range 0 to 32767
Remarks Refer to the Programming Guide for more information.